beautypg.com

Internal electronic status leds, Cpu status indicator, Figure 10-2: example of signal i/o function – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 242

background image

Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

224

Toggles parameter

ON/OFF

Exit returns to

DIAG display & all values

return to software control

DIAG SIGNAL I/O

PREV

NEXT

ENTR

EXIT

DIAG I/O

0 ) EXT_ZERO_CAL=ON

PREV

NEXT

JUMP

PRNT EXIT

DIAG I/O

22

)

WHEEL_HTR=

ON

PREV NEXT JUMP

ON

PRNT EXIT

DIAG I/O 28) SAMPLE_PRESSURE=2540 MV

PREV

NEXT

JUMP

PRNT EXIT

DIAG I/O

22

)

WHEEL_HTR=

OFF

PREV NEXT JUMP

OFF

PRNT EXIT

If parameter is an

input signal

If parameter is an output

signal or control

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

Figure 10-2:

Example of Signal I/O Function

10.1.4. Internal Electronic Status LEDs

Several LED’s are located inside the instrument to assist in determining if the analyzer’s CPU, I

2

C bus

and relay board, GFC wheel and the sync/demodulator board are functioning properly.

10.1.4.1. CPU Status Indicator

DS5, a red LED, that is located on upper portion of the motherboard, just to the right of the CPU board,
flashes when the CPU is running the main program loop. After power-up, approximately 30 to 60