Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

53

 Remove the set screw located in the top, center of the rear panel
 Remove the screws fastening the top cover to the unit (four per side).
 Slide the cover back and lift straight up .

4. Disconnect the current loop option PCA from the appropriate connector on the motherboard (see

Figure 5-1).

5. Place a shunt between the leftmost two pins of the connector (see Figure 5-1).

6. Reattach the top case to the analyzer.

7. The analyzer is now ready to have a voltage-sensing, recording device attached to that output


5.3. Expendable Kits (Options 42C, 42D and 43)

OPTION NUMBER

DESCRIPTION

OPT 42C

1 year’s supply of replacement of 47mm dia. particulate filters

OPT 42D

1 full replacement’s volume of indicating soda-lime for the external CO

2

scrubber

included with options 51 & 53 (approximate active lifetime: 1 year)

OPT 43

Options 42 C & 42D