Theory of Operation

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

204

The four sample and hold circuits are:

DESIGNATION ACTIVE

WHEN:

IR BEAM PASSING THROUGH

SEGMENT SENSOR PULSE IS:

Measure Gate

MEASUREMENT cell of GFC Wheel

HIGH

Measure Dark Gate

MEASUREMENT cell of GFC Wheel

LOW

Reference Gate

REFERENCE cell of GFC Wheel

HIGH

Reference Dark Gate

REFERENCE cell of GFC Wheel

LOW

Timing for activating the Sample and Hold circuits is provided by a phase lock loop circuit (PLL). Using
the segment sensor output as a reference signal the PLL generates clock signal at ten times that
frequency. This faster clock signal is used by the PLD to make the sample and hold circuits capture the
signal during the center portions of the detected waveform, ignore the rising and falling edges of the
detector signal.

Sample & Hold

Active

Detector

Output

Sample & Hold

Inactive

Figure 9-14:

Sample & Hold Timing

9.3.4.3. Sync/Demod Status LED’s

The following two status LED’s located on the synch/demod board provide additional diagnostic tools for
checking the GFC wheel rotation.

Table 9-1: Sync/Demod Status LED Activity

LED

FUNCTION

STATUS OK

FAULT STATUS

D1

M/R Sensor Status

LED flashes approximately

2/second

LED is stuck

ON or OFF

D2

Segment Sensor

Status

LED flashes approximately

6/second

LED is stuck

ON or OFF

See Section 11.1.4 for more information.