Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
85
6.7.2.8. RS-232 Report Function
The GFC 7000TA DAS can automatically report data to the communications ports, where they can be
captured with a terminal emulation program or simply viewed by the user.
To enable automatic COM port reporting, follow the instruction shown in Section 6.7.2.2 then press:
Edit Data Channel Menu
SETUP X.X
0) CONC: ATIMER, 1, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
SET> EDIT PRINT EXIT SETUP X.X RS-232 REPORT: OFF EDIT PRINT EXIT ENTR accepts the new setting and returns to the previous menu. EXIT ignores the new setting and returns to the previous menu. SETUP X.X RS-232 REPORT: OFF OFF ENTR EXIT Toggle to turn reporting ON or OFF Press SET> key until… From the DATA ACQUISITION menu (see Section 6.7.2.2) 6.7.2.9. Compact Report When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting 6.7.2.10. Starting Date This option allows to specify a starting date for any given channel in case the user wants to start data
each parameter in one channel on a separate line, up to five parameters are reported in one line.
acquisition only after a certain time and date. If the Starting Date is in the past, the DAS ignores this
setting.