6.11.5. COM Port Communication Modes

Each of the analyzer’s serial ports can be configured to operate in a number of different modes, listed in
Table 6-10 which can be combined by adding the mode ID numbers. For example, quiet mode, computer
mode and internet-enabled mode would carry a combined mode ID of 11, the standard configuration on
the GFC 7000TA COM2 port. Note that each COM port needs to be configured independently.

Table 6-10: Com Port Communication Modes

MODE

1

ID

DESCRIPTION

QUIET

1

Quiet mode suppresses any feedback from the analyzer (DAS reports, and warning
messages) to the remote device and is typically used when the port is communicating
with a computer program such as APICOM. Such feedback is still available but a
command must be issued to receive them.

COMPUTER

2

Computer mode inhibits echoing of typed characters and is used when the port is
communicating with a computer program, such as APICOM.

SECURITY

4

When enabled, the serial port requires a password before it will respond. The only
command that is active is the help screen (? CR).

HESSEN

PROTOCOL

16

The Hessen communications protocol is used in some European countries. Teledyne
Instruments part number 02252 contains more information on this protocol.

E, 7, 1

2048

When turned on this mode switches the com port settings
from

No parity; 8 data bits; 1 stop bit

to

Even parity; 7 data bits; 1 stop bit

RS-485

1024

Configures the COM2 Port for RS-485 communication. RS-485 mode has precedence
over multidrop mode if both are enabled.

MULTIDROP

PROTOCOL

32

Multidrop protocol allows a multi-instrument configuration on a single communications
channel. Multidrop requires the use of instrument IDs.

ENABLE

MODEM

64

Enables to send a modem initialization string at power-up. Asserts certain lines in the
RS-232 port to enable the modem to communicate.

ERROR

CHECKING

2

128

Fixes certain types of parity errors at certain Hessen protocol installations.

XON/XOFF

HANDSHAKE

2

256

Disables XON/XOFF data flow control also known as software handshaking.

HARDWARE

HANDSHAKE

8

Enables CTS/RTS style hardwired transmission handshaking. This style of data
transmission handshaking is commonly used with modems or terminal emulation
protocols as well as by Teledyne Instrument’s APICOM software.

HARDWARE

FIFO

2

512

Improves data transfer rate when on of the com ports.

COMMAND

PROMPT

4096

Enables a command prompt when in terminal mode.

1

Modes are listed in the order in which they appear in the

SETUP

MORE  com  COM[1 OR 2] MODE menu

2

The default sting for this feature is ON. Do not disable unless instructed to by Teledyne Instruments Customer Service

personnel.