Com port password security – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 169

background image

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

151

To initialize the modem press:


SETUP X.X

SECONDARY SETUP MENU

COMM

VARS DIAG ALRM

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SETUP X.X

COMMUNICATIONS MENU


ID INET

COM1

COM2

EXIT

EXIT

returns to

the previous

menu

SETUP X.X

COM1 MODEM INIT:AT Y

 &D &H


SET>

EDIT

EXIT

SETUP X.X

COM1 INITIALIZE MODEM

SET> INIT

EXIT

SETUP X.X

COM1 MODE:0

SET>

EDIT

EXIT

SETUP X.X COM1 BAUD RATE:19200

SET>

EDIT

EXIT

EXIT

returns to the

Communications Menu.

SETUP X.X

INITIALIZING MODEM


SET> INIT

EXIT

6.15.2.7. COM Port Password Security

In order to provide security for remote access of the GFC 7000TA, a LOGON feature can be enabled to
require a password before the instrument will accept commands. This is done by turning on the
SECURITY MODE

(Mode 4, Section 6.11.5). Once the SECURITY MODE is enabled, the following items

apply.

 A password is required before the port will respond or pass on commands.
 If the port is inactive for one hour, it will automatically logoff, which can also be achieved with the

LOGOFF command.

 Three unsuccessful attempts to log on with an incorrect password will cause subsequent logins to be

disabled for 1 hour, even if the correct password is used.

 If not logged on, the only active command is the '?' request for the help screen.
 The following messages will be returned at logon:

 LOGON SUCCESSFUL - Correct password given
 LOGON FAILED - Password not given or incorrect
 LOGOFF SUCCESSFUL - Connection terminated successfully

To log on to the Model GFC 7000TA analyzer with SECURITY MODE feature enabled, type:

LOGON 940331

940331 is the default password. To change the default password, use the variable RS232_PASS issued
as follows:

V RS232_PASS=NNNNNN