Setting the hessen protocol response mode, Text – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 174

Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
156
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG ALRM
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
COMMUNICATIONS MENU
ID INET
HESN
COM1
COM2
EXIT
SETUP X.
HESSEN VARIATION: TYPE 1
SET>
EDIT
EXIT
SETUP X.X
HESSEN VARIATION: TYPE 1
TYE1 TYPE 2
ENTR EXIT
SETUP X.X
HESSEN VARIATION: TYPE 2
PREV NEXT
OFF
ENTR EXIT
Press to change
protocol type.
ENTR
accepts the new
settings
EXIT
ignores the new
settings
NOTE
While Hessen Protocol Mode can be activated independently for COM1 and COM2,
The TYPE selection affects both Ports.
6.15.4.5. Setting The Hessen Protocol Response Mode
The Teledyne Instruments implementation of Hessen Protocol allows the user to choose one of several
different modes of response for the analyzer.
Table 6-30: Teledyne Instruments Hessen Protocol Response Modes
MODE ID
MODE DESCRIPTION
CMD
This is the Default Setting. Reponses from the instrument are encoded as the traditional command
format. Style and format of responses depend on exact coding of the initiating command.
BCC
Responses from the instrument are always delimited with
of the command encoding.
TEXT
Responses from the instrument are always delimited with
string, regardless of the command encoding.