beautypg.com

Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 199

background image

Maintenance

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

181

Table 8-1: GFC 7000TA Maintenance Schedule

ITEM ACTION FREQ

CAL

CHECK

REQ’D.

MANUAL

DATE PERFORMED

Particulate

Filter

Replace

Weekly or as

needed

Yes

Verify Test

Functions

Record and

analyze

Weekly or after

any

Maintenance

or Repair

No

Pump

Diaphragm

Replace Annually Yes

Perform

Flow Check

Check Flow

Annually

No

Perform

Leak Check

Verify Leak

Tight

Annually or

after any

Maintenance

or Repair

No

Pneumatic

lines

Examine

and clean

As needed

Yes if

cleaned

Cleaning Clean As

needed

Only if

cover

remv’d