Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

92

6.8.3. Single Range Mode (SNGL)

This is the default reporting range mode for the analyzer. In single range mode both A1 and A2 are set to
the same reporting range. This reporting range can be any value between 50 ppm and 2 000 ppm.

While the two outputs always have the same reporting range, the span, signal offset and scaling of their
electronic signals may be configured for differently (e.g., A1 = 0-10 V; A2 = 0-0.1 V). See Section 6.13.4
for instructions on adjusting these parameters.

To select SNGL range mode and to set the upper limit of the range, press:



SETUP C.3

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP C.3

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP C.3

RANGE MODE:

SNGL

SNGL

DUAL AUTO

ENTR EXIT

EXIT

x 2 returns

to the main

SAMPLE

display

SETUP C.3

RANGE MODE:

SNGL

SNGL

DUAL AUTO

ENTR

EXIT

SETUP C.3

RANGE CONTROL MENU

MODE

SET

UNIT

EXIT

SETUP C.3

RANGE:

500.0 Conc

0 0

5

0 0 .0

ENTR

EXIT

SETUP C.3

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SAMPLE* RANGE = 500.000 PPM CO2 X.XXX

< TST TST > CAL

SETUP