Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

68

To view and clear warning messages:

SAMPLE

SYSTEM RESET CO2 = 0.00

TEST

CAL

MSG

CLR

SETUP

Press CLR to clear the current

message.

If more than one warning is active, the

next message will take its place

Once the last warning has been

cleared, the analyzer returns to

SAMPLE

mode

SAMPLE

RANGE=500.000 PPM

CO2 = 0.00


< TST TST > CAL

MSG

CLR

SETUP

SAMPLE

SYSTEM RESET

CO2 = 0.00


TEST CAL

MSG

CLR

SETUP

TEST

deactivates warning

messages

MSG

displays warning messages.

NOTE:

If the warning message persists

after several attempts to clear it,

the message may indicate a

real problem and not an artifact

of the warm-up period

Figure 6-3:

Viewing and Clearing GFC 7000TA WARNING Messages

6.3. Calibration Mode

Pressing the CAL key switches the GFC 7000TA into multi-point calibration mode. In this mode, the user
can calibrate the instrument or check the instruments calibration with the use of calibrated zero or span
gases.

If the instrument includes either the zero/span valve option or IZS option, the display will also include
CALZ

and CALS keys. Pressing either of these keys also puts the instrument into multipoint calibration

mode.

 The

CALZ

key is used to initiate a calibration of the zero point.

 The

CALS

key is used to calibrate the span point of the analyzer. It is recommended that this span

calibration is performed at 90% of full scale of the analyzer’s currently selected reporting range.

Because of their critical importance and complexity, calibration operations are described in detail in
Section 7 of this manual. For more information concerning the zero/span, zero/span and IZS valve
options, See Section 5.4.