Initial o, Sensor calibration procedure, Calibration setup – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 58: Calibration method

background image

Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

40

3.8.1. Initial O

2

Sensor Calibration Procedure

If your instrument has an O

2

sensor option installed that should be calibrated as well.

3.8.1.1. O

2

Calibration Setup

The pneumatic connections for calibrating are as follows:

C

a

li

bra

ted N

2

at

HI

GH

Span

Co

nc

ent

rat

ion

C

a

lib

ra

te

d

O

2

at

20.

8%

Sp

an

C

on

cen

tr

at

io

n

VE
NT

Figure 3-16:

O2 Sensor Calibration Set Up

O

2

SENSOR ZERO GAS: Teledyne Instruments’ recommends using pure N

2

when calibration the zero

point of your O

2

sensor option.

O

2

SENSOR SPAN GAS: Teledyne Instruments’ recommends using 21% O

2

in N

2

when calibration the

span point of your O

2

sensor option.

3.8.1.2. O

2

Calibration Method

STEP 1 – SET O

2

SPAN GAS CONCENTRATION :

Set the expected O

2

span gas concentration.

This should be equal to the percent concentration of the O

2

span gas of the selected reporting range

(default factory setting = 20.8%; the approximate O

2

content of ambient air).