beautypg.com

Flow calibration – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 157

background image

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

139

6.13.8. Flow Calibration

The flow calibration allows the user to adjust the values of the sample flow rates as they are displayed on
the front panel and reported through COM ports to match the actual flow rate measured at the sample
inlet. This does not change the hardware measurement of the flow sensors, only the software calculated
values.

To carry out this adjustment, connect an external, sufficiently accurate flow meter to the sample inlet (see
Section 11 for more details). Once the flow meter is attached and is measuring actual gas flow, press:


Adjust these values

until the displayed

flow rate equals the

flow rate being

measured by the

independent flow

meter.

SAMPLE RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG ACAL DAS RNGE PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU


COMM VARS

DIAG

EXIT

Exit at

any time
to return

to main

the

SETUP

menu

DIAG

SIGNAL I / O

NEXT

ENTR EXIT

Exit returns

to the

previous menu

Repeat Pressing NEXT until . .

.

DIAG

FLOW CALIBRATION

PREV NEXT

ENTR EXIT

DIAG FCAL

ACTUAL FLOW: 654 CC / M


0 6 5 4 ENTR EXIT

ENTR accepts the

new value and

returns to the

previous menu

EXIT

ignores the

new value and

returns to the

previous menu

Adjust these values until

the displayed flow rate

equals the flow rate being

measured by the

independent flow meter.

SETUP X.X

ENTER DIAG PASS: 818


8 1 8

ENTR

EXIT