Theory of Operation

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

191

IR shining through both cells is effected
equally by interfering gas in the Sample

Chamber

M/R

is Shifted

Figure 9-5:

Effects of Interfering Gas on CO2 MEAS & CO2 REF

Thus, the difference in the peak heights and the resulting M/R ratio is only due to CO

2

and not to

interfering gases. In this way, Gas filter correlation rejects the effects of interfering gases and so that the
analyzer responds only to the presence of CO

2

.

To improve the signal-to-noise performance of the IR photo-detector, the GFC wheel also incorporates an
optical mask that chops the IR beam into alternating pulses of light and dark at six times the frequency of
the measure/reference signal. This limits the detection bandwidth helping to reject interfering signals
from outside this bandwidth improving the signal to noise ratio.

The IR Signal as the Photo-Detector sees it

after being chopped by the GFC Wheel

Screen

CO

2

REF

CO

2

MEAS

Figure 9-6:

Chopped IR Signal