Initial calibration procedure – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 55
Getting Started
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
37
3.8. Initial Calibration Procedure
The next task is to calibrate the analyzer.
To perform the following calibration you must have sources for zero air and span gas available for input
into the sample port on the back of the analyzer. See Section 3.1.3 for instructions for connecting these
gas sources.
While it is possible to perform this procedure with any range setting we recommend that you perform this
initial checkout using the 500 ppm range.
NOTE
The following procedure assumes that the instrument does not have any of the available
Zero/Span Valve Options installed.
See Section 7.4 for instructions for calibrating instruments possessing Z/S valve options.
1. Set the Analog Output Range:
SETUP X.X
CFG DAS
RNGE
PASS CLK MORE
EXIT
SAMPLE RANGE = 500.000 PPM
CO2
=X.XXX
< TST TST > CAL
SETUP
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
Press this button to set
the analyzer for
SNGL
DUAL
or AUTO ranges
SETUP X.X
RANGE: 500.000 CONC
0 0
5
0 0 .
0
ENTR EXIT
To change the value of the
reporting range span, enter the
number by pressing the key under
each digit until the expected value
appears.
Press this button to select the
concentration units of measure:
PPB, PPM, UGM, MGM
EXIT
ignores the new setting and
returns to the RANGE CONTROL
MENU
.
ENTR
accepts the new setting and
returns to the
RANGE CONTROL MENU
.
SETUP X.X
RANGE:
500.000 Conc
0 0 0 5 0 .0
ENTR
EXIT