Zero/span calibration on auto range or dual ranges – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 190

Calibration Procedures
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
172
7.5.1. Zero/Span Calibration on Auto Range or Dual Ranges
If the analyzer is being operated in dual range mode or auto range mode, then the high and low ranges
must be independently calibrated.
When the analyzer is in either dual or auto range modes the user must run a separate calibration
procedure for each range. After pressing the CAL, CALZ or CALS buttons the user is prompted for the
range that is to be calibrated as seen in the CALZ example below:
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST
TST
>
CAL
CALZ
CALS SETUP
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
Wait until STABIL
falls below 1.0 ppm.
This may take several
minutes.
SAMPLE* RANGE = 500.000 PPM CO2 =XXX.X
TST> CAL CALZ CALS SETUP SAMPLE STABIL=XXX.X PPM CO2 =XXX.X < TST TST > CAL CALZ CALS SETUP Set the Display to show the STABIL test function. This function calculates the stability of the CO 2 measurement Continue Calibration as per Standard Procedure ZERO CAL M RANGE = 500.000 PPM CO2 =XXX.X < TST TST > ZERO SPAN CONC EXIT ANALYZER ENTERS ZERO CAL MODE SAMPLE RANGE TO CAL: LOW LOW HIGH ENTR SETUP SAMPLE RANGE TO CAL: HIGH LOW HIGH ENTR SETUP Once this selection is made, the calibration procedure continues as previously described in Section 7.2.
The other range may be calibrated by starting over from the main SAMPLE display.