Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

170

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST >

ENTR

CONC

EXIT

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST

TST

>

CAL

CALZ

CALS SETUP

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST >

ZERO

CONC

EXIT

EXIT

returns to the main

SAMPLE

display

ACTION:

Allow zero gas to enter the sample port at the

rear of the instrument.

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST >

SPAN

CONC

EXIT

M-P CAL RANGE = 500.000 PPM CO2 =XXX.X

< TST TST >

ENTR

SPAN CONC

EXIT

M-P CAL RANGE = 500.000 PPM CO2 =XXX.X

< TST TST > ENTR CONC

EXIT

Wait until STABIL

falls below1.0 ppm.

This may take several

minutes.

The SPAN button now

appears during the

transition from zero to

span.

If either the ZERO or

SPAN

buttons fail to

appear see Section 11

for troubleshooting tips.

SAMPLE* RANGE = 500.000 PPM CO2 =XXX.X

TST>

CAL

CALZ CALS SETUP

SAMPLE

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST >

CAL

CALZ CALS SETUP

Set the Display to show the

STABIL

test function.

This function calculates the

stability of the CO

x

measurement

Press ENTR to changes the

OFFSET

& SLOPE values for the

CO

2

measurements.

Press EXIT to leave the calibration

unchanged and return to the

previous menu.

ACTION:

Allow span gas to enter the sample port at the

rear of the instrument.

The value of

STABIL

may jump

significantly.

Wait until it falls back

below 1.0 ppm.

This may take several

minutes.

Press ENTR to change the

OFFSET

& SLOPE values for the

CO

2

measurements.

Press EXIT to leave the calibration

unchanged and return to the

previous menu.