Calibration Procedures
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
170
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
ENTR
CONC
EXIT
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST
TST
>
CAL
CALZ
CALS SETUP
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
ZERO
CONC
EXIT
EXIT
returns to the main
SAMPLE
display
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
SPAN
CONC
EXIT
M-P CAL RANGE = 500.000 PPM CO2 =XXX.X
< TST TST >
ENTR
SPAN CONC
EXIT
M-P CAL RANGE = 500.000 PPM CO2 =XXX.X
< TST TST > ENTR CONC
EXIT
Wait until STABIL
falls below1.0 ppm.
This may take several
minutes.
The SPAN button now
appears during the
transition from zero to
span.
If either the ZERO or
SPAN
buttons fail to
appear see Section 11
for troubleshooting tips.
SAMPLE* RANGE = 500.000 PPM CO2 =XXX.X
TST> CAL CALZ CALS SETUP SAMPLE STABIL=XXX.X PPM CO2 =XXX.X < TST TST > CAL CALZ CALS SETUP Set the Display to show the STABIL test function. This function calculates the stability of the CO x measurement Press ENTR to changes the OFFSET & SLOPE values for the CO 2 measurements. Press EXIT to leave the calibration unchanged and return to the previous menu. ACTION: Allow span gas to enter the sample port at the rear of the instrument. The value of STABIL may jump significantly. Wait until it falls back below 1.0 ppm. This may take several minutes. Press ENTR to change the OFFSET & SLOPE values for the CO 2 measurements. Press EXIT to leave the calibration unchanged and return to the previous menu.