11.4.2. Basic anti-ESD Procedures for Analyzer Repair and
Maintenance
11.4.2.1. Working at the Instrument Rack
When working on the analyzer while it is in the instrument rack and plugged into a properly grounded
power supply.
1. Attach your anti-ESD wrist strap to ground before doing anything else.
Use a wrist strap terminated with an alligator clip and attach it to a bare metal portion of the
instrument chassis. This will safely connect you to the same ground level to which the
instrument and all of its components are connected.
2. Pause for a second or two to allow any static charges to bleed away.
3. Open the casing of the analyzer and begin work. Up to this point, the closed metal casing of your
analyzer has isolated the components and assemblies inside from any conducted or induced
static charges.
4. If you must remove a component from the instrument, do not lay it down on a non-ESD
preventative surface where static charges may lie in wait.
5. Only disconnect your wrist strap after you have finished work and closed the case of the analyzer.
11.4.2.2. Working at an Anti-ESD Work Bench.
When working on an instrument of an electronic assembly while it is resting on an anti-ESD work bench:
1. Plug your anti-ESD wrist strap into the grounded receptacle of the work station before touching
any items on the work station and while standing at least a foot or so away. This will allow any
charges you are carrying to bleed away through the ground connection of the workstation and
prevent discharges due to field effects and induction from occurring.
2. Pause for a second or two to allow any static charges to bleed away.
3. Only open any anti-ESD storage bins or bags containing sensitive devices or assemblies after
you have plugged your wrist strap into the workstation.
Lay the bag or bin on the workbench surface.
Before opening the container, wait several seconds for any static charges on the outside
surface of the container to be bled away by the workstation’s grounded protective mat.
4. Do not pick up tools that may be carrying static charges while also touching or holding an ESD
Sensitive Device.
Only lay tools or ESD-sensitive devices and assemblies on the conductive surface of your
workstation. Never lay them down on any non-ESD preventative surface.
5. Place any static sensitive devices or assemblies in anti-static storage bags or bins and close the
bag or bin before unplugging your wrist strap.
6. Disconnecting your wrist strap is always the last action taken before leaving the workbench.