beautypg.com

Manual calibration checks – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 183

background image

Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

165

7.3. Manual Calibration Checks

Informal calibration checks, which only evaluate but do not alter the analyzer’s response curve, are
recommended as a regular maintenance item and in order to monitor the analyzer’s performance. To
carry out a calibration check rather than a full calibration, follow these steps.

STEP ONE

: Connect the sources of zero air and span gas as shown in Figures 7.1 or 7.2.

STEP TWO

: Perform the zero/span calibration check procedure:

SAMPLE

STABIL=XXX.X PPM

CO2=X.XXX


< TST TST >

CAL

SETUP

ACTION:

Record the CO

2

concentration

reading.

ACTION:

Supply span gas to the instrument

SAMPLE

STABIL=XXX.X PPM

CO2=X.XXX

< TST TST >

CAL

SETUP

ACTION:

Record the CO

2

concentration

reading.

Wait until

STABIL

is

below 1.0 ppm.

This may take

several minutes.

The value of

STABIL

may jump

significantly.

Wait until it falls

below 1.0 ppm. This

may take several

minutes.

SAMPLE

STABIL=XXX.X PPM

CO2=X.XXX

< TST TST >

CAL

SETUP

The SPAN key appears during the transition from zero to

span. You may see both SPAN and ZERO buttons.

SAMPLE RANGE = 500.0 PPM

CO2=X.XXX

< TST TST >

CAL

SETUP

Scroll the display to the

STABIL

test function.

ACTION:

Supply the instrument with zero gas.