Manual calibration checks – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 183
Calibration Procedures
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
165
7.3. Manual Calibration Checks
Informal calibration checks, which only evaluate but do not alter the analyzer’s response curve, are
recommended as a regular maintenance item and in order to monitor the analyzer’s performance. To
carry out a calibration check rather than a full calibration, follow these steps.
STEP ONE
: Connect the sources of zero air and span gas as shown in Figures 7.1 or 7.2.
STEP TWO
: Perform the zero/span calibration check procedure:
SAMPLE
STABIL=XXX.X PPM
CO2=X.XXX
< TST TST >
CAL
SETUP
ACTION:
Record the CO
2
concentration
reading.
ACTION:
Supply span gas to the instrument
SAMPLE
STABIL=XXX.X PPM
CO2=X.XXX
< TST TST >
CAL
SETUP
ACTION:
Record the CO
2
concentration
reading.
Wait until
STABIL
is
below 1.0 ppm.
This may take
several minutes.
The value of
STABIL
may jump
significantly.
Wait until it falls
below 1.0 ppm. This
may take several
minutes.
SAMPLE
STABIL=XXX.X PPM
CO2=X.XXX
< TST TST >
CAL
SETUP
The SPAN key appears during the transition from zero to
span. You may see both SPAN and ZERO buttons.
SAMPLE RANGE = 500.0 PPM
CO2=X.XXX
< TST TST >
CAL
SETUP
Scroll the display to the
STABIL
test function.
ACTION:
Supply the instrument with zero gas.