Page 182

Calibration Procedures
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
164
STEP THREE: Perform the Zero/Span Calibration Procedure:
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
ENTR
CONC
EXIT
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
CAL
SETUP
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
ZERO
CONC
EXIT
EXIT
returns to the main
SAMPLE
display
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
SPAN
CONC
EXIT
M-P CAL RANGE = 500.000 PPM CO2 =XXX.X
< TST TST >
ENTR
SPAN CONC
EXIT
M-P CAL RANGE = 500.000 PPM CO2 =XXX.X
< TST TST > ENTR CONC
EXIT
Wait until STABIL
falls below 1.0 ppm.
This may take several
minutes.
The SPAN button
appears.
If either the ZERO or
SPAN
buttons fail to
appear see Section 11
for troubleshooting tips.
SAMPLE* RANGE = 500.000 PPM CO2 =XXX.X
< TST
TST
>
CAL
SETUP
SAMPLE
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
CAL
SETUP
Set the Display to show the
STABIL
test function.
This function calculates the
stability of the CO
2
measurement
Press ENTR to changes the
OFFSET
& SLOPE values for the
CO
2
measurements.
Press EXIT to leave the calibration
unchanged and return to the
previous menu.
ACTION:
Allow span gas to enter the sample port at the
rear of the instrument.
The value of
STABIL
may jump
significantly.
Wait until it falls back
below 1.0 ppm
This may take several
minutes.
Press ENTR to change the
OFFSET
& SLOPE values for the
CO
2
measurements.
Press EXIT to leave the calibration
unchanged and return to the
previous menu.
If the ZERO or SPAN buttons are not displayed, this means that the measurement made during that part
of the procedure is too far out of the allowable range to do allow a reliable calibration. The reason for this
must be determined before the analyzer can be calibrated. See Section 11 for troubleshooting tips.