140
6.13.9. Test Channel Output
When activated, output channel A4 can be used to report one of the test functions viewable from the
SAMPLE mode display. To activate the A4 channel and select a test function, follow this button
sequence:
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
DIAG
ANALOG OUTPUT
PREV
NEXT
ENTR
EXIT
DIAG TCHN
TEST CHANNEL: NONE
NEXT
ENTR
EXIT
DIAG
TEST CHAN OUTPUT
PREV
NEXT
ENTR
EXIT
Continue to press NEXT until …
DIAG TCHN
TEST CHANNEL: CO2 MEASURE
PREV
NEXT
ENTR
EXIT
Press PREV or NEXT
to move through the
list of available
parameters
(Table 6-13)
Press EXIT to
return to the
DIAG
menu
Press ENTR to select
the displayed
parameter activating
the test channel.
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
ALRM
EXIT
Table 6-22: Test Parameters Available for Analog Output A4
TEST CHANNEL
ZERO
FULL SCALE
NONE
Test Channel is turned off
CO
2
MEASURE
0 mV
5000 mV*
CO
2
REFERENCE
0 mV
5000 mV*
SAMPLE PRESS
0 "Hg
40 "Hg
SAMPLE FLOW
0 cc/m
1000 cc/m
SAMPLE TEMP
0
C 70C
BENCH TEMP
0
C 70C
WHEEL TEMP
0
C 70C
CHASSIS TEMP
0
C 70C
PHT DRIVE
0 mV
5000 mV