Calibration quality – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 194

Calibration Procedures
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
176
7.6.1. AutoCal with Auto or Dual Reporting Ranges Modes Selected
SETUP C.4
RANGE TO CAL: HIGH
EDIT EXIT EXIT returns to the PRIMARY SETUP Menu SETUP C.4 SEQ 2) ZERO–SPAN, 2:00:30 MODE SET EXIT SETUP C.4 RANGE TO CAL: LOW LOW HIGH ENTR SETUP SETUP C.4 RANGE TO CAL: LOW EDIT EXIT NOTE In order to automatically calibrate both the HIGH and LOW ranges, you must set up a separate sequence for each. 7.7. Calibration Quality After completing one of the calibration procedures described above, it is important to evaluate the Make sure that these parameters are within the limits listed in Table 7-3 and frequently compare them to Table 7-3: Calibration Data Quality Evaluation FUNCTION MINIMUM VALUE OPTIMUM VALUE MAXIMUM VALUE SLOPE 0.700 1.000 1.300 OFFS -0.500 0.000 0.500 These values should not be significantly different from the values recorded on the Teledyne Instruments sheet that was shipped with your instrument. If they are, refer to the troubleshooting Section 11.
PREV NEXT
analyzer’s calibration SLOPE and OFFSET parameters. These values describe the linear response curve
of the analyzer. The values for these terms, both individually and relative to each other, indicate the
quality of the calibration. To perform this quality evaluation, you will need to record the values of both test
functions (Section 6.2.1 or Appendix A-3), all of which are automatically stored in the DAS channel
CALDAT for data analysis, documentation and archival.
those values on the Final Test and Checkout Sheet that came attached to your manual, which should not
be significantly different. If they are, refer to the troubleshooting Section 11.
Final Test and Validation Data