Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

78

6.7.2.3. Editing DAS Data Channels

DAS configuration is most conveniently done through the APICOM remote control program. The
following list of key strokes shows how to edit using the front panel.

Main Data Acquisition Menu

Edit Data Channel Menu

SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX


< TST TST > CAL

SETUP

SETUP X.X

0) CONC: ATIMER, 1, 900

PREV NEXT

INS DEL EDIT PRNT EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG

DAS

RNGE

PASS CLK MORE

EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

EXIT

will return to the

previous SAMPLE

display.

SETUP X.X

DATA ACQUISITION

VIEW

EDIT

EXIT

Moves the

display up &

down the list of

Data Channels

Inserts a new Data

Channel into the list

BEFORE the Channel

currently being displayed

Deletes The Data

Channel currently

being displayed

Exports the

configuration of all

data channels to

RS-232 interface.

Exits to the Main

Data Acquisition

Menu

SETUP X.X

NAME:CONC

EDIT PRNT

EXIT

Moves the display

between the

PROPERTIES for this

data channel.

Reports the configuration of current

data channels to the RS-232 ports.

Exits returns to the

previous Menu

Allows to edit the channel name, see next key sequence.

When editing the data channels, the top line of the display indicates some of the configuration
parameters. For example, the display line:

0) CONC: ATIMER, 4, 800

translates to the following configuration:
Channel No.: 0
NAME: CONC
TRIGGER EVENT: ATIMER
PARAMETERS: Four parameters are included in this channel
EVENT: This channel is set up to record 800 data points.