6.13.4.5. AIN Calibration
This is the sub-menu to conduct the analog input calibration. This calibration should only be necessary
after major repair such as a replacement of CPU, motherboard or power supplies. Activate the ANALOG
I/O CONFIGURATION MENU (See Section 6.13.1), then press:
Exit at any time to
return to the main
DIAG
menu
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
Continue pressing SET? until …
DIAG AIO
AIN CALIBRATED:
NO
< SET
SET>
CAL
EXIT
DIAG AIO
AIN CALIBRATED:
YES
< SET
SET> CAL
EXIT
Instrument
calibrates
automatically
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
STARTING FROM ANALOG I / O CONFIGURATION MENU
DIAG AIO
CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
6.13.4.6. Configuring Analog Inputs (Option) Channels
To configure the analyzer’s external analog inputs option, define for each channel:
gain (number of units represented by 1 volt)
offset (volts)
engineering units to be represented in volts (each press of the touchscreen button
scrolls the list of alphanumeric characters from A-Z and 0-9)
whether to display the channel in the Test functions
To access and adjust settings for the external Analog Inputs option channels press: