Dark calibration test, Meas and on for co – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 155
Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
137
6.13.6. Dark Calibration Test
The dark calibration test interrupts the signal path between the IR photo-detector and the remainder of
the sync/demod board circuitry. This allows the instrument to compensate for any voltage levels inherent
in the sync/demod circuitry that might effect the calculation of CO
2
concentration. Performing this
calibration returns two offset voltages, One for CO
2
MEAS
and on for CO
2
REF
that are automatically
added to the CPU’s calculation routine. The two offset voltages from the last calibration procedure may
be reviewed by the user via the front panel display.
To activate the dark calibration procedure or review the results of a previous calibration, press:
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
ALRM
EXIT
Calibration runs automatically
Electric offset for Measurement signal
Electric offset for Reference signal
DIAG
SIGNAL
I
/
O
NEXT
ENTR
EXIT
Repeat Pressing NEXT until . . .
DIAG DARK
REF DARK OFFSET: 0.0 mV
EXIT
DIAG DARK
MEAS DARK OFFSET: 0.0 mV
EXIT
DIAG DARK
DARK CAL 1% COMPLETE
EXIT
DIAG DARK
DARK CALIBRATION ABORTED
EXIT
Display
tracks %
complete
DIAG
DARK CALIBRATION
PREV NEXT
ENTR
EXIT
Exit returns
to the
previous menu
DIAG DARK
CO2 DARK CALIBRATION
VIEW CAL
EXIT