Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

137

6.13.6. Dark Calibration Test

The dark calibration test interrupts the signal path between the IR photo-detector and the remainder of
the sync/demod board circuitry. This allows the instrument to compensate for any voltage levels inherent
in the sync/demod circuitry that might effect the calculation of CO

2

concentration. Performing this

calibration returns two offset voltages, One for CO

2

MEAS

and on for CO

2

REF

that are automatically

added to the CPU’s calculation routine. The two offset voltages from the last calibration procedure may
be reviewed by the user via the front panel display.

To activate the dark calibration procedure or review the results of a previous calibration, press:


SETUP X.X

ENTER DIAG PASS: 818


8 1 8

ENTR

EXIT

SAMPLE RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

ALRM

EXIT

Calibration runs automatically

Electric offset for Measurement signal

Electric offset for Reference signal

DIAG

SIGNAL

I

/

O

NEXT

ENTR

EXIT

Repeat Pressing NEXT until . . .

DIAG DARK

REF DARK OFFSET: 0.0 mV

EXIT

DIAG DARK

MEAS DARK OFFSET: 0.0 mV

EXIT

DIAG DARK

DARK CAL 1% COMPLETE

EXIT

DIAG DARK

DARK CALIBRATION ABORTED

EXIT

Display

tracks %

complete

DIAG

DARK CALIBRATION

PREV NEXT

ENTR

EXIT

Exit returns

to the

previous menu

DIAG DARK

CO2 DARK CALIBRATION

VIEW CAL

EXIT