beautypg.com

Half-bridge configurations – Measurement Computing WaveBook rev.3.0 User Manual

Page 61

background image

WaveBook User’s Manual,

ch03B 6-23-99

WBK Expansion Options, WBK16 3-29

Half-Bridge Configurations
The half-bridge has two strain-variable elements and requires two internal bridge completion resistors
(BCRs). Any bridge resistance from 60 to 1000 ohms can be accommodated for either the internal or
external bridge.

Half-Bridge (+), Any Resistance from 60 to 1000 Ohms,

B,D, or F Shunt

IA

External
Bridge

DB-9

Input

Internal Bridge Completion

RA

RB

RC

RD

RE

RF

RG

RH

1

2

3

-Sense

+Sense
+Excitation

-Excitation

JN01

Switches

accessed

through

software

1

8

9

7

5

4

3

6

2

In this connection, excitation voltage is
regulated at the connector. This
configuration should only be used for
short cable lengths
. Output polarity can
be altered by selecting the software invert
function. The B, D, or F shunt resistor
may be software selected. Output polarity
during shunt calibration will be
automatically corrected by software.

IA

External

Bridge

DB-9
Input

Internal Bridge Completion

RA

RB

RC

RD

RE

RF

RG

RH

1

2

3

-Sense

+Sense
+Excitation

-Excitation

JN01

Switches

accessed

through

software

Half-Bridge (+), with Remote Sense

1

8

9

7

5

4

3

6

2

In this connection, excitation voltage is
regulated at the strain gage. This is the
preferred connection for longer cables.

3-Wire TC Half-Bridge, Software Invert & B, D, F Shunt Available

IA

External
Bridge

DB-9

Input

Internal Bridge Completion

RA

RB

RC

RD

RE

RF

RG

RH

1

2

3

-Sense

+Sense
+Excitation

-Excitation

JN01

Switches

accessed

through

software

1

8

9

7

5

4

3

6

2

This occasionally utilized connection can
be made as shown. Two resistors
normally reserved for quarter-bridge
completion must be used. For
compatibility with other configurations,
the above two configurations are
preferred instead of this one.