Acquisition Operation

Acquisition Modes

The acquisition mode determines the make-up of the acquisition—the number of scans and their relation to
the trigger. The 4 acquisition modes are: N-shot, infinite post-trigger, N-shot with re-arm, and pre/post-
trigger.

N-Shot Acquisition Mode

The N-shot mode, shown below, is the simplest. After the trigger, it collects a fixed number of scans,
from 1 to 100,000,000.

Scan 1

Scan 2

Scan 3

Scan N

Scan N-1

No acquisitions
before trigger

Scan Period

Samples in Scan

Trigger

Time

N-Shot Acquisition

1 µs/sample

Channel

ch1

ch3

ch5

ch2

ch7

ch1

×1

×5

×10

×5

×1

×5

Gain

Uni

Uni

Uni

Bi

Bi

Bi

Offset

After the acquisition has started with the daqArm command, the N-shot mode waits until a trigger occurs.
After the trigger, it collects the post-trigger number of scans (as specified by wbkSetAcq) and then stops.
No more data will be collected until another acquisition has been started. As shown above, each scan
consists of one or more samples, taken 1 microsecond apart. The scans are repeated with the post-trigger
scan period that was specified with daqAdcSetRate or daqAdcSetFreq. The number of scans
collected (N) may be set from a minimum of one scan after the trigger to a maximum of 100,000,000 scans.

Infinite Post-Trigger Acquisition Mode

It is possible to collect an unlimited number of scans after the trigger by using the infinite post-trigger
acquisition mode as shown below.

Scan Period

Infinite Post-Trigger Acquisition

Trigger

Disarm

Time

This mode is similar to the N-shot mode except the total number of scans is not limited. With infinite post-
trigger, scans are continually taken at the post-trigger scan rate. When the desired number of scans have
been gathered, the daqDisarm command must be used to terminate acquisition.

N-Shot with Re-Arm Acquisition Mode

The N-shot with re-arm acquisition mode is used to collect scans after each trigger, as shown below.

Scan 1

Scan N

No acquisitions

before trigger

No acquisitions

after dis-arm

Scan Period

N-Shot With Re-Arm Acquisition

Trigger

Time

Trigger

Trigger

Disarm

****

This mode is similar to the N-shot mode except that after the "N" scans have been acquired, the trigger is
re-armed so another "N" scans will be taken for each subsequent trigger. As in the infinite post-trigger
mode, the daqDisarm command must be invoked to prevent further acquisitions.