Measurement Computing WaveBook rev.3.0 User Manual

Page 104

background image

5-10 WaveView,

6-24-99

WaveBook User’s Manual

Units. This column allows you to change the voltage scale setting of each analog channel displayed

when the Scope option is selected. When a cell is selected, a selection box gives you a choice between
volts (V) or millivolts (mV). You can also enter user units and mX+b scales from this point. Making a
selection sets the choice into the individual cell or block of cells. This option has no effect on the
Digital Input channel.

Auto-Zero (32-Bit WaveView Only). Auto-zero is used to null-out any DC offset that might be present

in a channel. WaveView nulls-out the offsets of all channels set to “Yes” in the Auto-Zero column,
providing the channels are enabled (“On”).

The following options depend on your actual equipment configuration.

LPF On (WBK12/13, WBK14, WBK16). This column allows you to include or exclude the low-pass

filter from a channel. When selecting a cell, the selection box above the spreadsheet allows “On” or
“Bypass” to enable or disable the filter or block of filters. In WBK14 applications, LPF On allows you
to select an external filter. Double-clicking a cell in the LPF On column will cycles through the
available options.

LPF Cutoff (WBK12/13, WBK14, WBK16). This column allows you to set the low-pass filter cut-off

frequency for the selected channel(s). If you enter an inappropriate cut-off frequency, the software will
round up or down to the next appropriate frequency for your particular hardware. Since the WBK12/13
filters are assigned to banks, setting one channel of a bank will also update the others.

LPF Type (WBK12/13, WBK14, WBK16). This column allows you to configure the low-pass filter for

the selected channel(s). When selecting a cell or block of cells in this column, a selection box above the
spreadsheet may or may not appear, depending upon your particular hardware. If the selection box
appears, it will display the appropriate low-pass filter selections (such as “Elliptic” or “Linear” for the
WBK12/13) allowed by your hardware. Double-clicking a cell in this column will toggle the filter-type
status. A change in the low-pass filter type for one channel will appropriately update any other affected
channels.

HPF Cutoff (WBK14, WBK16). This column allows you to set the high-pass filter cut-off frequency

for the selected channel(s). When a cell is selected, a selection box above the spreadsheet will display
the appropriate cut-off frequency selections (such as “0.1 Hz” or “10 Hz” for a WBK14 or “DC” and
“10 Hz” for a WBK16) to configure the filter(s). Double-clicking a cell in this column will toggle the
cut-off frequency status. A change in the high-pass filter cut-off frequency for one channel will
appropriately update any other affected channels.

Source Level (WBK14, WBK16). This column allows you to apply or remove the source level for the

selected channel(s). When selecting a cell or block of cells in this column, a selection box above the
spreadsheet may or may not appear, depending upon your particular hardware. If the selection box
appears, it will display the appropriate source level selections (such as “Off”, “2 mA”, or “4 mA” for a
WBK14 or an excitation voltage level for a WBK16) allowed by your hardware to configure a source
or block of sources. Double-clicking a cell in this column will toggle the source level status. A change
in the source level for one channel will appropriately update any other affected channels.
Note: For WBK14 Only, when using an ICP transducer, either 2 mA or 4 mA must be selected.
Set the current-source level to “Off” before measuring voltage.

Bridge Type (WBK16 Only). This column allows you to select the specific bridge configuration for a

strain gage or load cell sensor. When a cell is selected, a selection box above the spreadsheet will
display the appropriate bridge configuration selections (such as Full Bridge, Half-Bridge and
Quarter-Bridge). Double-clicking a cell in this column toggles the Bridge Type.

Invert (WBK16 Only). This column allows you to invert the signal level of a channel. When a cell is

selected, the selection box above the spreadsheet allows “Yes” or “No” options to determine whether
the channel is inverted. Double-clicking a cell in this column will toggle the invert status.