Hypertherm HyIntensity Fiber Laser Rev.3 User Manual
Page 128

setup and OperatiOn
3-20
Hypertherm Fiber Laser
Instruction Manual – 807090 Revision 3
Setups field
value
Combiner Parameters
Combiner Quad
Combiner Slope
Combiner Offset
Values are set at
the factory
Hypernet
ON when using a Hypertherm CNC
OFF when using any other CNC
The user selects
the value
Comb Test
ON = Combiner Stray light Sensor is active
OFF = Combiner Stray light Sensor is not active
The user selects the
value (Default = ON)
MPA Display
ON = Gas pressure value is converted to MPA
OFF = Gas pressure value is shown in PSI
The user selects
the value
PWM Display
ON = PWM Duty displayed on main screen
OFF = PWM Duty is not displayed
The user selects
the value
+
=
Setups
screen
Continued from the previous page
Push to Select
+
=
Combiner
Parameter
screen
Push to Select
Caution: the values on the
factory and should not be changed without instructions from
the factory or from a Hypertherm technician.
+
=
+
=
Navigate to other
setup fields and set
parameters to the
values in the table
Hypernet
OFF
MPA Display
OFF
Comb Test
ON
PWM Display
OFF
Comb Quad
x.x
Comb Slope
x.x
Comb Offset
x.x
Hypernet
OFF
MPA Display
OFF
Comb Test
ON
PWM Display
OFF
Password 15
Press Select to accept new values
Timeout in 5 seconds