Test Setup - Test Configurator, Timer, and System

Packet Blazer

47

Test Configurator Overview

Physical Interface Port indicates the physical interface port.

To ensure proper optical module detection/validation, make sure that
the CFP/CFP2 screws are fully secured.

The status for a physical CFP/CFP2 interface and the compatibility with
the selected signal/interface is indicated as follows. The icon and its
description are also displayed just below the test application name
when validating or if there is a CFP/CFP2 problem.

For serial CFP interface, an interlink fault status indicates a problem at
the interconnect between the serial CFP and the module. No icon is
displayed when there is no interlink fault. The icon and its description
is also displayed just below the test application name when there is a
CFP problem.

Icon

Description

Validating CFP/CFP2

Missing CFP/CFP2

Invalid CFP/CFP2 or mismatch with the selected signal/interface.

Valid CFP/CFP2 matching the selected interface/rate.

Icon

Description

Fault detected on either TX, RX, or TX/RX interlink. If an interlink fault is
reported we recommend to try the following to fix the issue: Fully remove the
CFP, wait 30 seconds then fully re-insert and secure the CFP with its screws.