5.3.2 Details on Parameter Setting Commands
The following describes how to change the BL-700’s settings through command
communication.
You can use the setup software to change the BL-700’s settings instead of these
commands.
Communication Details
1. Send the direct control command SSET to the BL-700.
The BL-700 will shift to setting mode.
After successfully executing the command, the BL-700 sends back an OK.
2. After the BL-700 shifts to setting mode, send the command (setting change
command) corresponding to the item to be changed to the BL-700.
After successfully executing the command, the BL-700 sends back an OK.
If an error occurs with this command, the BL-700 sends back ERRnn. (nn
sttands for error code numbers.)
3. To confirm the current settings, send a setting confirmation command.
After successfully executing the command, the BL-700 sends back an OK.
If an error occurs with this command, the BL-700 sends back ERRnn.
4. To save the current settings in the EEP-ROM, send SAVE to the BL-700.
Once you save the settings in the EEP-ROM, the BL-700 will start with these
settings when the BL-700 is turned on next.
After successfully executing the command, the BL-700 sends back an OK.
If an error occurs with this command, the BL-700 sends back ERRnn.
5. To quit the setting mode and perform the normal bar code reading, send SEND
to the BL-700.
After successfully executing the command, the BL-700 sends back an OK.
If an error occurs with this command, the BL-700 sends back ERRnn.
Communication format
When the command format is [Command][CR], the response format is
[Response][CR]. When the command format is [STX][Command][ETX], the
response format is [STX][Response][ETX].
•
•
When
command receiving buffer are cleared.
•
If the BL-700’s command receiving buffer contains erroneous characters due to a data
transmission error during communication, add
Note: For command communication, set the time duration between transmission of
each character (byte) to up to 30 seconds. If this duration exceeds 30 seconds, the
BL-700 cancels the received characters.