beautypg.com

Quantification data tab, Quantification tab spreadsheet, Results spreadsheet – Bio-Rad Firmware & Software Updates User Manual

Page 82

background image

MiniOpticon Instruction Manual

71

Quantification Tab Spreadsheet

Table 24 shows the type of data shown in the spreadsheet at the bottom right side of the
Quantification tab:

TIP: To make changes to the Content, Target, and Sample, open the Plate Editor by
clicking the Plate Setup button and selecting View/Edit Plate.

Quantification Data Tab

The Quantification Data tab shows spreadsheets that describe the quantification data
collected in each well. Select one of the four options to show the data in different formats:

• Results. Displays a spreadsheet view of the data

Standard Curve Results. Displays a spreadsheet view of the standard curve data

• Plate. Displays a view of the data in each well as a plate map

• RFU. Choose this spreadsheet to show the RFU quantities in each well for each cycle

TIP: Right-click any spreadsheet for options, including the sort option.

Results Spreadsheet

Select a Results spreadsheet (Figure 51) to see data for each well in the plate.

Figure 51. Quantification Data tab with Results spreadsheet selected.

Baseline Thresholds...

Open the Baseline Thresholds window to change baseline
or thresholds of each fluorophore (changes appear in
Amplification chart in Quantification tab).

Table 24. Quantification tab spreadsheet content

Information

Description

Well

Well position in the plate.

Fluor

Fluorophore detected.

Target

Target Name loaded in the Plate Editor wells.

Content

A combination of the Sample Type (required) and Replicate #
(optional) loaded in the Plate Editor.

Sample

Sample Name loaded in the Plate Editor wells.

C

q

Quantification cycle for each trace.

Table 23. Right-click menu items for spreadsheets. (continued)

Menu Option

Function