Calibration Wizard

The MiniOpticon system is factory calibrated for commonly used fluorophores in Bio-Rad
white-welled plates (Table 43).

To open the Calibration Wizard to calibrate the MiniOpticon system:

1. Select an instrument in the Detected Instruments pane.

2. Select Tools > Dye Calibration Wizard to open the Dye Calibration window and

calibrate new dye and plate combinations (Figure 85).

Figure 85 shows an example of the Dye Calibration window.

Figure 85. Dye Calibration window.

Calibrating the MiniOpticon System

To calibrate the MiniOpticon system in the Dye Calibration window:

1. Select the Plate Type in the Calibrate New or Existing Fluorophores pane. If the plate

type is not included in the list, type the name in the box to add it to the list.

2. Select the fluorophore you want to calibrate from the pull-down list. If the fluorophore

name is not included in the list, type the name in the box to add it to the list.

3. Begin preparing a 48-well plate for dye calibration by pipetting dye solution into each

well. For each fluorophore, fill each well with 50 µl of 300 nM dye solution.

4. Seal the plate using the sealing method you will use in your experiment.

5. Click Next and follow the on-screen instructions. First, place an empty plate in the

MiniOpticon system for the first procedure of the calibration. Next, place the calibration

Table 43. Factory calibrated fluorophores, channels, and instruments.

Fluorophores

Channel

FAM, SYBR

®

Green I

1

HEX

2