Custom export tab – Bio-Rad Firmware & Software Updates User Manual
Page 132
MiniOpticon Instruction Manual
121
• NRT (no reverse transcriptase control) with a C
q
less than XX. Input a C
q
Cut-off
Value
• Positive control with a C
q
greater than XX. Input a Cq Cut-off Value
• Unknown without a C
q
• Standard without a C
q
• Efficiency greater than XX. Input a reaction efficiency Cut-off Value that is calculated
for the standard curve
• Efficiency less than XX. Input a reaction efficiency Cut-off Value that is calculated for
the standard curve
• Std Curve R^2 less than XX. Input a Cut-off R^2 value for the standard curve
• Replicate group C
q
Std Dev greater than XX. Input a Cut-off standard deviation that
is calculated for each replicate group
Custom Export Tab
Select the Custom Export tab (Figure 82) to define the default settings for the fields that will
be exported and their export format when the Custom Export option is chosen.
Figure 82. Custom Export tab in the User Preferences window.
The File export formats include text (*.txt), CSV (*.csv), Excel 2007 (*.xlsx), Excel 2003 (*.xls),
XML (*.xml), and HTML (*.html).
The following items can be chosen for export:
• Sample Description. Well, Fluorophore, Target Name, Content, Replicate Number,
Sample Name, Biological Set Name, and Well Note
• Quantification. C
q
, Starting Quantity, C
q
Standard Deviation, and Quantity Standard
Deviation
• Melt Curve. Melt Temperature, Peak Height, Melt Peak Begin Temperature, and End
Temperature
• End Point. End Point Call and End RFU