Select fluorophores window – Bio-Rad Firmware & Software Updates User Manual
Page 55
Plates
44
6. Click OK to apply the information entered to your plate. Any additional edits can be
made using the Plate Editor.
NOTE: Returning to the Auto layout tab will reset the plate layout. The layout can
also be reset by selecting Tools > Clear Plate.
TIP: To read text within wells of the plate layout more easily, click the Zoom plate
box for a magnified view.
Select Fluorophores Window
The Select Fluorophores window lists fluorophores that can be selected to load into the Plate
Editor well loading controls. To open the Select Fluorophores window, click the Select
Fluorophores button on the right side of the Plate Editor.
NOTE: You cannot add or remove fluorophores in this list; you must calibrate the
new fluorophores on an instrument in the Calibration Wizard (page 126). After
calibration, the new fluorophore is added to the Select Fluorophore window.
NOTE: The MiniOpticon system is factory calibrated only for FAM, SYBR
®
and
HEX. If you intend to use another dye, you must perform a dye calibration. To
calibrate a new combination of dye and plate type on an instrument, select Tools >
Dye Calibration Wizard (see “Calibration Wizard” on page 126).
Click the Selected check box next to the fluorophore name to add or remove the fluorophores
to the list on the right side of the Plate Editor window.
In this example, SYBR
®
is selected from the list of available fluorophores (Figure 30).
Figure 30. Select Fluorophores window.
• Click the Color box next to the fluorophore name and select a new color to represent
each fluorophore in the Plate Editor window and Data Analysis charts
NOTE: Before beginning the run, the software verifies that the fluorophores you
specified in the plate are calibrated on that instrument. You cannot run a plate if it
includes fluorophores that have not been calibrated on that instrument.