beautypg.com

Protocol autowriter, Creating a protocol with the protocol autowriter – Bio-Rad Firmware & Software Updates User Manual

Page 48

background image

MiniOpticon Instruction Manual

37

• Calculated mode. When you enter a sample volume between 1 and 50 µl the thermal

cycler calculates the sample temperature based on the sample volume. This is the
standard mode

• Block mode. When you enter a sample volume of zero (0) µl, the thermal cycler records

the sample temperature as the same as the measured block temperature

Protocol AutoWriter

Open the Protocol AutoWriter to quickly write protocols for PCR and real-time PCR runs. To
open the Protocol AutoWriter, select one of these options:

• Click the Protocol AutoWriter button in the main software window toolbar

• Select Tools > Protocol AutoWriter from the menu bar in the main software window

Figure 27 shows a protocol (bottom of window) written by the Protocol AutoWriter.

Figure 27. Protocol AutoWriter window with a new protocol.

Creating a Protocol with the Protocol AutoWriter

Follow these steps to use the Protocol AutoWriter to create a new protocol:

1. Click the Protocol AutoWriter button on the toolbar to open the Protocol AutoWriter

window.

2. Enter the Annealing Temperature (Ta) and Amplicon Length in the boxes within the

Enter Target Values/Enzymes pane. If you do not know the annealing temperature for
primers, click the Ta Calculator button to enter the primer sequences and calculate the
annealing temperature. For information about the calculations used in the Ta Calculator
see Breslauer et al. 1986.