Sensor Installation and Connection 3-31
Leveling the Surface Combo Sensor or Ultrasonic Depth Sensor
If necessary, adjust the level for either sensor in the following
way:
Remove the sensor from the mounting plate.
Loosen the ring slightly to allow the plate to move on the
ring.
Tap the sensor mounting plate to the right or left with a
rubber mallet until it is level.
Warning: Tapping the mounting plate with the sensor
attached could damage the sensor. Always remove the
sensor before tapping the plate with a mallet.
Reattach the sensor to the mounting plate, and recheck the
level.