8-50 IETG FlowHawk Manual
Problem
Depth data indicates a surcharged pipe, but the pipe is
free flowing.
Possible
Causes
Physical offset may be incorrect.
Electronic offset may be incorrect.
Pipe height may be incorrect.
Sensor may be dirty.
Sensor may be faulty.
Possible
Solutions
Verify the physical offset.
Verify the pipe height.
Verify electronic offset and adjust if necessary.
Clean the sensor.
Contact your regional IETG representative.
Replace the sensor.
Replace the monitor if defective.
Problem
Ultrasonic depth readings are erratic.
Possible
Causes
Noise may exist in the sewer pipe.
Sensor may be dirty.
Flow may be choppy or foamy.
Sensor may not be level.
Possible
Solutions
Check the flow conditions.
Clean the sensor.
Check if the sensor is level and correct if necessary.