Modbus/local (host) port configuration – ADS Environmental Services IETG FlowHawk QR 775012 A4 User Manual
Page 372

D-4
IETG FlowHawk Manual
• The system will exit the current menu or submenu following 30
seconds of inactivity. To return to the initial location before
the timeout occurred, enter three consecutive bangs (!!!) to
return to the main menu from the READY state or hit the
appropriate number key to return to a particular submenu from
the main menu.
• Exit any submenu by entering a lower case “x”.
The following sections provide instructions on how to complete the
parameters or diagnostics in each submenu.
Modbus/Local (Host) Port Configuration
Selecting this submenu allows you to configure the local (host) port
and corresponding modes:
• Baud Rate Select the number (1-3) corresponding to the
desired baud rate. An asterisk (*) will display beside the
selected option. The new baud rate will be initiated after
restarting the EMUX, which occurs through the System Tests
menu.
• Intermessage timeout This parameter currently is not
configurable by the user. It defaults to 2.
• Modbus Communications Mode Enter 5 to toggle
between the two available modes. The initial in parentheses (B
or P) indicates the current mode.
Buffered Choosing this mode maintains the set of real-
time data from the monitor on the EMUX. The unit
provides this data upon request from the RTU. This data is
updated in the EMUX based on the scan rate in the
monitor.
Passthrough Choosing this mode causes all data
requests from the RTU to go directly to the monitor.