8-14 IETG FlowHawk Manual
Carefully slide the thick red band securing the PCB on the
side of the battery pack off of the PCB.
Carefully disconnect the PCB from the battery pack at the
white plastic connector.
Disconnecting the PCB from the battery pack
Set the PCB temporarily aside in a clean, dry location. Do
not discard! This PCB will be used with the replacement
battery pack.
11. Connect the PCB to the replacement battery pack and then
secure the PCB to the side of the pack with the red band.
12. Run the loose cable from the battery pack through the hole in
the underside of the retainer plug, and then seat the retainer
plug on top of the battery pack. Be sure to pull through the
excess cable before fully seating the plug.
13. Lower the battery pack into the monitor canister, and then
connect the cable from the battery pack to the monitor
top/dome assembly.