General monitor problems – ADS Environmental Services IETG FlowHawk QR 775012 A4 User Manual

Page 323

background image

Maintenance and Troubleshooting 8-39

Some problems that occur will not require a site visit, such as
incorrect equipment identification numbers or other system
parameters the user can re-enter on the local PC. However, some
problems will require a site visit. When this is necessary, inform
the data analyst any time a field crew is en route to a monitor site to
troubleshoot problems so that the analyst can attempt to collect the
monitor data before they arrive. If the problem is a faulty monitor
and the analyst cannot collect the data remotely, replace the monitor
and deliver the faulty monitor to the office so the analyst can
attempt to collect the data directly. Then, send the monitor to IETG
for repair.

This section provides general guidelines for troubleshooting and
correcting problems with the FlowHawk monitor and sensor
subsystems.

General Monitor Problems

The following tables contain general troubleshooting techniques for
the FlowHawk flow monitor and communications.

Problem

Time stamp on the collected data is incorrect.

Possible

Causes

PC clock may have been set to incorrect time when the
monitor was activated.

Monitor time zone difference for the PC clock may not
have been set correctly when the monitor was activated.

Monitor clock may be faulty.

Possible

Solutions

Verify the time on the PC clock and correct if necessary.

Reactivate the monitor to enable the clock.

Verify the time zone setting of the monitor. Correct and
reactivate if necessary.

Collect the data from the monitor and replace monitor if
defective.