beautypg.com

Figure 4-15 program action screen -33, Figure 4-16 select defined action screen -33 – INFICON MDC-260 Thin Film Deposition Controller User Manual

Page 69

background image

MDC-260 DEPOSITION CONTROLLER

PROGRAMMING AND CONTROLLER SETUP

4-33

Once you have selected the required action, pressing the Right-arrow key will
present the screen which permits programming of the action details, and this
procedure is similar to the one used for programming discrete outputs.

Figure 4-15 Program Action Screen

In this screen you select the predefined action you would like to take and the
associated conditions. To specify an action, move the markers onto the action
name field and press the Right-arrow key. This will present the Select Defined
Action screen.

Figure 4-16 Select Defined Action Screen

In this screen you can select a predefined action from the list by moving the
cursors onto the desired action and pressing Enter. The following is a list of the
predefined actions:

No Action - No action is taken. This is the default setting.

Manual - Functionally identical to pressing Manual key.

Zero - Functionally identical to pressing Zero key.

Reset - Functionally identical to pressing Reset key.

Abort - Functionally identical to pressing Abort key.

Halt - Halts the process, sets active source power to idle, and leaves all other
source powers unchanged.

Terminate Deposit - Triggers the final thickness for the deposit state. Action is
ignored if state is not a deposit state.

Hold In State - Holds controller in current state.

Action Name

Hold In State

-----

Conditions





Press > to select action

Select Defined Action: 01

No Action

-------

02 Manual Power

03 Zero Thickness

04

Reset

Controller

05

Abort

Process

06

Halt

Process

07

TerminateDeposit

to select action

08 Hold In State