beautypg.com

Code #28: send controller status, Code #28: send controller status -36 – INFICON MDC-260 Thin Film Deposition Controller User Manual

Page 174

background image

MDC-260 DEPOSITION CONTROLLER

COMPUTER INTERFACE

11-36

Code #28: Send Controller Status

Description:

Instructs the controller to send the controller status data list.


Format/Example:

• To instruct the controller to send the controller status data list, transmit:

Header

Interface

Address

Instruction

Code

Length

Checksum

Dec

255 254

1

28 0

228

Hex

FF FE

01

1C 00 E4


Response:

• In addition to the normal transmission receipt, the controller will respond

with:

Header

Interface

Address

Inst.

Code

Length

24 Data

Bytes

Checksum

Dec

255 254

1

28

24

(data) (checksum)

Hex

FF FE

01

1C

18

(data) (checksum)

(data):

The 24 data bytes represent the controller status data list.

(checksum):

The checksum from the instruction code through the last data byte.


A description of the controller status data list is as follows:

Parameter Name

Length

(bytes)

Values

Controller State

1

1 = Process ready
2 = Start layer
3 = Change pocket
4 = Change crystal
5 = Layer ready
6 = Soak rise
7 = Soak hold
8 = Rise to Predeposit
9 = Predeposit hold
10 = Establish rate
11 = Shutter Delay
12 = Deposit #1
13 = Rate ramp #1
14 = Deposit #2
15 = Rate ramp #2