Code #25: receive output setup, Code #26: send action setup, Code #25: receive output setup -33 – INFICON MDC-260 Thin Film Deposition Controller User Manual
Page 171: Code #26: send action setup -33

MDC-260 DEPOSITION CONTROLLER
COMPUTER INTERFACE 11-33
Code #25: Receive Output Setup
Description:
Instructs the controller to enter the incoming parameters for Output
#N into memory. Use the table from Code #24 to construct a message to set the
parameters for Output #N to your desired settings.
Format/Example:
• To re-program Output #N with custom parameters, transmit:
Header
Interface
Address
Inst.
Code
Length
Output #N
(1-8)
43 Data
Bytes
Checksum
Dec
255 254
1
22
44
2 (data)
(checksum)
Hex
FF FE
01
16
2C
02 (data)
(checksum)
(data):
The 43 data bytes represent the configuration parameters the selected
Output.
(checksum):
The checksum from the instruction code through the last data byte.
Response:
• The controller will respond with the normal transmission receipt.
Code #26: Send Action Setup
Description:
Instructs the controller to send the parameter list for Action #N. A
description of the Action parameter list is as follows:
Parameter Name
Length (bytes)
Allowable Range
Action Type
1
0...37
Condition String
24
! | & ( )
Total: 25 Data Bytes
Format/Example:
• To instruct the controller to send the parameter list for Action #5, transmit:
Header
Interface
Address
Instruction
Code
Length
Action #N
(1-8)
Checksum
Dec
255 254
1
26 1 5 223
Hex
FF FE
01
1A 01 05 DF