beautypg.com

Saving and restoring user settings, Inputs, Saving and restoring user settings inputs – Watlow EZ-ZONE RMH User Manual

Page 104: Calibration offset, Calibration

background image

Watlow EZ-ZONE

®

RMH Module

101

Chapter 6 Features

Saving and Restoring User Settings

Recording setup and operations parameter settings

for future reference is very important. If you uninten-

tionally change these, you will need to program the

correct settings back into the controller to return the

equipment to operational condition.

After you program the controller and verify prop-

er operation, use User Save Set [USr;S] (Setup Page,

Global Menu) to save the settings into either of two

files in a special section of memory. If the settings

in the controller are altered and you want to return

the controller to the saved values, use User Restore

Set [USr;r] (Setup Page, Global Menu) to recall one

of the saved settings. A digital input or the Function

Key can also be configured to restore parameters.

Note:

Starting with firmware release 6, there is only one

user set.

Note:

Only perform the above procedure when you are

sure that all the correct settings are programmed

into the controller. Saving the settings overwrites

any previously saved collection of settings.

Be sure to document all the controller settings.

Note:

When restoring factory defaults, I/O assemblies for

Modbus, DeviceNet, Profibus and Ethernet along

with the zone address will be overwritten when re-

storing factory defaults.

Inputs

Calibration Offset

Calibration offset allows a device to compensate for

an inaccurate sensor, lead resistance or other factors

that affect the input value. A positive offset increases

the input value, and a negative offset decreases the

input value.

The input offset value can be viewed or changed

with Calibration Offset [`i;CA] (Operations Page,

Analog Input Menu).

Time

Temperature

Temperature Reading

from Sensor

Actual Process Temperature

Negative Calibration Offset will

compensate for the difference

between the Sensor Reading and

the Actual Temperature

Calibration

To calibrate an analog input, you will need to provide

two electrical signals or resistance loads near the ex-

tremes of the range that the application is likely to

utilize. See recommended values below:

Sensor Type

Low Source

High Source

thermocouple

0.000 mV

50.000 mV

millivolts

0.000 mV

50.000 mV

volts

0.000V

10.000V

milliamps

0.000 mA

20.000 mA

100 Ω RTD

50.00 Ω

350.00 Ω

1,000 Ω RTD

500.00 Ω

3,500.00 Ω

Thermistor 5K

50.00 Ω

5000.00 Ω

Thermistor 10K

50.00 Ω

10000.00 Ω

Thermistor 20K

50.00 Ω

20000.00 Ω

Thermistor 40K

50.00 Ω

40000.00 Ω

Follow these steps for a thermocouple or pro-

cess input:

1. Apply the low source signal to the input you are

calibrating. Measure the signal to ensure it is ac-

curate.

2. Read the value of Electrical Measurement [`Mu]

(Factory Page, Calibration Menu) for that input.

3. Calculate the offset value by subtracting this val-

ue from the low source signal.

4. Set Electrical Input Offset [ELi;o] (Factory Page,

Calibration Menu) for this input to the offset val-

ue.

5. Check the Electrical Measurement to see whether

it now matches the signal. If it doesn’t match, ad-

just Electrical Offset again.

6. Apply the high source signal to the input. Mea-

sure the signal to ensure it is accurate.

7. Read the value of Electrical Measurement for

that input.

8. Calculate the gain value by dividing the low

source signal by this value.

9. Set Electrical Slope [ELi;S] (Factory Page, Cali-

bration Menu) for this input to the calculated

gain value.

10. Check the Electrical Measurement to see wheth-

er it now matches the signal. If it doesn’t match,

adjust Electrical Slope again.

Set Electrical Offset to 0 and Electrical Slope to 1 to

restore factory calibration.
Follow these steps for an RTD input:

1. Measure the low source resistance to ensure it

is accurate. Connect the low source resistance to

the input you are calibrating.

2. Read the value of Electrical Measurement [`Mu]

(Factory Page, Calibration Menu) for that input.

3. Calculate the offset value by subtracting this val-

ue from the low source resistance.

4. Set Electrical Input Offset [ELi;o] (Factory Page,