Chapter 8 Calibration 8-15

Module: ADAM-4016

1. Apply power to the strain gauge input module and let it warm up for

about 30 minutes.

2. Assure that the module is correctly installed. Then, connect a

voltmeter to the screw terminals of the module.

Exc-

Voltmeter

-

+

Exc+

Ii

n

-

Ii

n

+

Vin-

Vin+

ADAM-4016

Figure 8-7 Setup for Voltage Output Calibration

3. Execute the Zero Calibration command. This is done through the

ADAM utility software. (Please see "A/O 0V Calibration" option in
the Calibration sub-menu of the ADAM utility software.)

Figure 8-8 Zero Calibration