Setting method (measured correction), Setting method (measured correction) -43, F'(.6 – KEYENCE IB Series User Manual
Page 95: 3 calibration function, Press the [set] button, Ib-e, Set ting v a rious f unct ions, Main screen, Measurement mode" (page 4-8)
4-3 Calibration Function
4-43
IB-E
4
Set
ting V
a
rious
F
unct
ions
Setting Method (Measured Correction)
Main screen
1
Press the [MODE] and buttons for approx. 2
seconds on the main screen at the same time.
[30.CLb] is displayed on the main display (upper level).
Press the [MODE]
+ buttons for 2
seconds or more
at the same time
2
Press the / button to set [MEAS] and press the
button.
set1
3
Insert workpiece A for which the measured value is
known, and set “R.V. you wish to display (1)” for the
workpiece A using the / buttons.
The current internal measurement value is displayed on the
main display (upper level). [set1] and the “R.V. you wish to
display (1)” are alternately displayed on the sub display's
screen (lower level).
4
Press the [SET] button.
After [set1] blinks several times on the sub display's screen
(lower level), [set2] is displayed.
[SET] button
set2
5
Insert workpiece B for which the measured value is
known, and set “R.V. you wish to display (2)” for the
workpiece B using the / buttons.
The current internal measurement value is displayed on the main
display (upper level). [set2] and the “R.V. you wish to display (2)”
are alternately displayed on the sub display's screen (lower level).
[SET] button
LASER
BANK
0
1
2
3
HI
GO
LO
R.V. ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
HOLD CALC CHECK
F'(.6
LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI
SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
%.D
GO
HOLD
CALC
CHECK
Calibration setting
Setting value
Description
Default value
defLt
No correction (Default setting)
MEAS
Measured correction
LoGic
Logical correction
LASER
BANK
0
1
2
3
HI
GO
LO
R.V. ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
HOLD CALC CHECK
Setting you wish to display
R.V.
" 1. Measurement Mode" (page 4-8)
Item
Measurement
mode
Setting range
Default
value
R.V. you
wish to
display (1)
% mode
- 999.99 to 999.99
0.00
Dimension mode
- 99.999 to 99.999
0.000
LASER
BANK
0
1
2
3
HI
GO
LO
R.V. ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
HOLD CALC CHECK
" 1. Measurement Mode" (page 4-8)
Item
Measurement
mode
Setting range
Default
value
R.V. you
wish to
display (2)
% mode
- 999.99 to 999.99
100.00
Dimension mode
- 99.999 to 99.999
10.000