3-2 Operations on the Main Screens

3-7

IB-E

3

B

a

sic O

perat

io

ns

Functions available on the setting screen

Main screen

Press the [MODE] button for approx. 2 seconds.

4-2 Basic Settings and Advanced Settings (page 4-4)

Basic settings
Basic settings, such as measurement mode and averaging,
are made.
Advanced settings
More advanced settings such as hold function, delay timer
enable the unit to be used in more complex applications.

Press the [MODE] and  buttons for approx. 2 seconds.

4-3 Calibration Function (page 4-42)

When there is a difference between the internal
measurement value (R.V.) or calculated value (CALC value)
and the actual dimension of the object, the value can be
calibrated.

While pressing the [MODE] button, press the [SET] button 5
times.

3-5 Initial Reset (Initialize) (page 3-12)

All settings, excluding the calibration function, are initialized.

Press the  and  buttons for approx. 2 seconds.

3-3 Beam Axis Adjustment (page 3-8)

This function assists adjusting the laser beam’s axis.

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT