Operations on the Main Screens

This section describes R.V. (Internal Measurement Value) displayed on the sub display
(lower level) and P.V. (Judgment Value) displayed on the main display (upper level).

R.V. (Internal Measurement Value)
R.V. (Internal Measurement Value) is the value displayed when a target is inserted into the
measurement range.
* R.V. = Raw Value

P.V. (Judgment Value)
P.V. (Judgment Value) is the value to set the judgment output to ON/OFF according to the
tolerance setting value. Also, the analog output is output based on the P.V.
* P.V. = Present Value

"3-6 Setting the Tolerance Setting Value" (page 3-13)

The judgment value (P.V.) and the internal measurement value (R.V.) are basically the
same, however, those values differ only when the hold function is used.

"5. Hold Function" (page 4-14)

The judgment value (P.V.) is displayed on the main display.
The display varies as below according to whether the hold function is used or not.

R.V. (Internal Measurement Value) and P.V. (Judgment Value)

Main Display (Upper Level)

Normal
The same value as the internal measurement value (R.V.) is displayed
as a judgment value (P.V.).

When the hold function is used
The judgment value (P.V.) is held according to the hold function setting.

"5. Hold Function" (page 4-14)