3-6 Setting the Tolerance Setting Value
3-15
IB-E
3
B
a
sic O
perat
io
ns
Tolerance tuning
When the target (master workpiece) as a reference is present, the HIGH setting value
(upper limit) and LOW setting value (lower limit) can automatically be set with the master
workpiece measurement value as the center value.
1
Press the / button several times on the main screen. Then display the
R.V. value display screen on the sub display (lower level).
"Sub Display (Lower Level)" (page 3-4)
2
Measure the master workpiece and press the [SET] button.
The judgment value (P.V.) as a reference value for the tolerance setting is imported.
[set] and the tolerance tuning setting width are displayed alternately on the sub
display (lower level).
3
Press the / button to set the tolerance tuning setting width.
4
Press the [SET] button to complete the tolerance tuning.
[set] blinks on the main display (upper level), and the HIGH setting value and LOW
setting value are determined. Then, the display returns to the R.V. screen
automatically.
Automatic Setting
Point
Tolerance tuning cannot be performed when the displayed P.V.
(judgment value) is [-----], [FFFF] or [-FFFF]. [no.uAL] will blink
several times on the main display if attempted.
LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI
SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
HOLD
CALC
CHECK
GO
R.V.
[R.V.]
ON
GO
HOLD
CALC
CHECK
LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI
SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
Tolerance tuning setting width
Master workpiece P.V.
Tolerance tuning
setting width
Measurement upper limit
Measurement lower limit
LOW side setting value (lower limit)
HIGH side setting value (upper limit)
Master workpiece P.V. (judgment value)
Item
Measurement mode
Setting range
Default value
Tolerance tuning
setting width
% mode
0.00 to 999.99
10.00
Dimension mode
0.000 to
99.999
1.000