3-6 Setting the Tolerance Setting Value
3-17
IB-E
3
B
a
sic O
perat
io
ns
5
Measure the good target again and press the [SET] button. (LOW side 1st
point confirmation operation)
The internal measurement value (R.V.) is imported as a good target measurement
value.
[Loset] is displayed on the main display (upper level).
6
Measure the LOW side defective target and press the [SET] button. (LOW
side 2nd point confirmation operation)
The internal measurement value (R.V.) is imported as a measurement value for LOW
side defective target.
After [set] blinks on the main display (upper level), the judgment value (P.V.) is
displayed.
On the sub display (lower level), the median value of the good target value (1st point)
imported on step 5 and LOW defective target value (2nd point) is displayed.
Setting the LOW setting value (lower limit) is complete.
The 2-point tuning is complete.
Measurement value of
HIGH limit defective target
Measurement value of
LOW limit defective target
Measurement value of acceptable target
Measurement upper limit
Measurement lower limit
LOW setting value (lower limit)
HIGH setting value (upper limit)